Standalone Calibration of Meters, Clamps and Current Transformers
The 2558A is a cost-effective, high-precision instrument capable of outputting a wide range of AC voltages and currents for the calibration of AC measurement devices such as analog wattmeters, power meters, clamp-on testers and current transformers. Intuitive front panel controls and various computer interfaces let the 2558A be operated manually or integrated into a larger system.
Key Features
- Voltage, current and frequency sweeping
- Output divider
- Direct readout of the deviation
- Digital display of output on high-visibility 7-segment LEDs
- Common current output terminals
- High accuracy and stability
- Wide output and frequency range
- Dedicated dials and switches for each digit and function
- Automation through communication command control
- Wide variety of external signal interfaces
Specifications
| AC voltage ranges | 100 mV / 1 V / 10 V / 100 V / 300 V / 1000 V – specified 1 mV to 1200.0 V |
| AC current ranges | 100 mA / 1 A / 10 A / 50 A – specified 1 mA to 60.00 A |
| Voltage accuracy | ±0.04% – ±(0.03% of setting + 0.01% of range) at 50/60 Hz on 100 mV–10 V |
| Current accuracy | ±0.05% – ±(0.04% of setting + 0.01% of range) at 50/60 Hz on 100 mA–1 A |
| Frequency range | 40 to 1000 Hz |
| Stability | ±(20 ppm of setting + 30 ppm of range) |
| Distortion factor | Voltage output 0.07% or less; current output 0.18% or less |
| Maximum output | Approx. 36 VA |
| Maximum output current | Approx. 3 A on the 10 V range; approx. 0.5 A on the 1 V range |