Accurate, Reliable Calibration of AC Power Measurement Devices
Built on Yokogawa’s history of developing superior sourcing instruments, the LS3300 is an AC signal generator capable of outputting both current and voltage simultaneously, allowing it to calibrate power measurement instruments such as current clamps and transformers, panel meters, RMS multimeters and power meters. The LS3300 delivers 78.5 kW per unit, with user-configurable current, voltage, frequency and phase settings. A high-definition LCD display, vector diagram and the ability to synchronize up to three units let it emulate anything from single-phase two-wire to three-phase four-wire systems.
Key Features
- Intuitive interface for quickly adjusting voltage, current, frequency and phase values
- High-definition LCD display for clean and clear data visualization
- Functions tailored to power measurement verification – signal ramping, stepping and triggering, plus synchronization for multi-phase power generation
- Vector display for tracking current/voltage phase relationships
- Harmonic output option superimposes a single harmonic component up to the 51st order
- Software to automatically verify economy-class power meters and generate calibration reports
- Independently source voltage and current sine waves into the kilohertz range with high accuracy and resolution
- High test uncertainty ratio with most economy-class power meters, making it a trusted reference for instrument verification and calibration
Specifications
| AC voltage output | 10 mV to 1250 V – ranges 1 / 10 / 30 / 100 / 300 / 1000 V |
| AC current output | 0.3 mA to 62.5 A – ranges 30 mA / 100 mA / 1 A / 10 A / 50 A |
| Paralleled current ranges | 20 A and 100 A with two units; 30 A and 150 A with three units |
| AUX output | 500 mV range (0 to 625.00 mV) and 5 V range (0 to 6.2500 V) |
| Voltage accuracy (1 year) | ±350 ppm – ±(0.03% of setting + 0.005% of range) at 50/60 Hz |
| Current accuracy (1 year) | ±450 ppm – ±(0.04% of setting + 0.005% of range) at 50/60 Hz |
| Power accuracy (1 year) | ±450 ppm |
| Phase accuracy | ±0.03° at 50/60 Hz |
| Stability | Voltage and current ±50 ppm/h; power ±100 ppm/h |
| Frequency range | 40 to 1200 Hz |
| Output power | 78.5 kW per unit |
| Wiring emulation | 1P2W single unit; 1P3W, 3P3W and 3P4W with up to three synchronized units |
| Harmonic output | Optional (/HRM1) – single component up to the 51st order |