Two Isolated Source-Measure Channels, Expandable to Ten
The GS820 Multi-Channel Source Meter comes in 18-volt and 50-volt models for testing LED lighting, measuring drain current and more. It has two isolated analog channels, each with a constant voltage source, constant current source, voltmeter and ammeter. The GS820 offers two sweep functions, PC connectivity via USB, and expansion up to 10 channels.
Key Features
- Isolated 2-channel source and measurement function
- Source and measurement ranges – 18 V model: 7 V and 3.2 A or 18 V and 1.2 A; 50 V model: 20 V and 1.2 A or 50 V and 0.6 A
- 1 pA resolution at the extremely small 200 nA current range
- Generate arbitrary waveforms consisting of up to 100,000 points at 100 µs intervals
- Channel expansion through master-slave synchronization link
- Fast test speeds
- 16-bit digital I/O on models 765602 and 765612
- Four-quadrant operation – source operation (current source) and sink operation (current sink)
- 5½-digit output and measurement resolution
- Two sweep functions and USB PC connectivity
Specifications
| Channels | 2 isolated analog channels, expandable to 10 by master-slave link |
| Operation | Four-quadrant source and sink |
| Resolution | 5½ digits, output and measurement |
| 18 V model – voltage ranges | 200 mV / 2 V / 7 V / 18 V (±200.000 mV to ±18.0000 V) |
| 18 V model – current ranges | 200 nA to 3 A (±200.000 nA to ±3.20000 A) |
| 50 V model – voltage ranges | 200 mV / 2 V / 20 V / 50 V (±200.000 mV to ±50.0000 V) |
| 50 V model – current ranges | 200 nA to 1.0 A (±200.000 nA to ±1.20000 A) |
| Voltage accuracy (1 year) | ±(0.02% of setting + 250 µV) on the 200 mV range |
| Current accuracy (1 year) | ±(0.06% of setting + 3 nA) on the 200 nA range |
| Arbitrary waveform | Up to 100,000 points at 100 µs intervals |
Orderable Models
| Model | Voltage range | Digital I/O |
|---|---|---|
| 765601 | 18 V | 2-bit |
| 765602 | 18 V | 16-bit |
| 765611 | 50 V | 2-bit |
| 765612 | 50 V | 16-bit |