A Multimeter with the Analysis of an Oscilloscope
The DM7560 is a flexible 6.5-digit digital multimeter with high accuracy and high-speed data logging at up to 30 kS/s, deep 100 k reading memory and built-in statistical analysis. Not a typical multimeter, it combines the benefits of an oscilloscope with the dedicated functionality of a DMM – measurements can be viewed as a trendline, histogram or arc scale with real-time statistical analysis.
Key Features
- Full-colour, high-resolution display with flexible display formats
- High-speed data logging with up to 30 k points per second
- High-capacity internal memory up to 100 k points
- Trend / histogram analysis available both in real time and offline
- Multiple PC interface options (USB, Ethernet, GP-IB, RS-232) enable automation
- Choose from several display formats: trendline, histogram or arc scale
- Built-in power line cycle (PLC) filter removes 50/60 Hz harmonics from measurements
- Limit and GO/NO-GO functionality for transient analysis
- Use standalone or as part of an integrated automated test system
- Delta-sigma A/D converter; NMRR 55 dB, CMRR 120 dB
Specifications
| Resolution | 6.5 digits |
| DC voltage ranges | 100 mV / 1 V / 10 V / 100 V / 1000 V |
| DC voltage accuracy | ±(0.0035% of reading + 0.0005% of range) on the 10 V range |
| DC current ranges | 1 mA / 10 mA / 100 mA / 1 A / 3 A |
| DC current accuracy | ±(0.050% of reading + 0.005% of range) on the 100 mA range |
| AC voltage ranges | 100 mV / 1 V / 10 V / 100 V / 750 V, true RMS |
| AC voltage bandwidth | 20 Hz to 300 kHz (20 Hz to 100 kHz on the 750 V range) |
| AC voltage accuracy | ±(0.06% of reading + 0.03% of range), 100 Hz to 20 kHz |
| AC current ranges | 1 A / 3 A, true RMS, 20 Hz to 5 kHz |
| Other functions | 2-wire and 4-wire resistance, temperature (thermocouple and RTD) |
| Sampling rate | Up to 30 kS/s on DC functions |
| Internal memory | Up to 100,000 readings |
| Interfaces | USB, Ethernet, GP-IB, RS-232 |
| Note | Internal reference junction compensation is not supported for thermocouples; the reference junction temperature is entered manually. |
Applications include battery current consumption monitoring, sensor testing, production testing, R&D and service work, and voltage reference testing.