GX7017 Series Specifications
Switching Subsystem Specifications | |
---|---|
GX6032 Switch Card | Configuration: 32 x16 Max Inputs: 32 to 128 (4 switch cards) |
GX6256 LF Switch Card | Configuration: 16 x16 matrix with 2 to 1 selector, (16) 2:16 multiplexers Inputs: 16 internal analog bus and 16 digital Outputs: 256 Maximum of 9 cards (2304 channels) per system |
GX6192 HF Switch Card | Configuration: 16 x 16 matrix with (16) 1:12 multiplexers Inputs: 16 external analog inputs or 16 internal analog bus Outputs: 192 |
GX6864 RF Switch Card | Bandwidth: 500 MHz Impedance: 75Ω Configuration: (4) 2 x 16 multiplexers Inputs: 16 internal analog bus, 8 external |
Relay Specifications | Max AC pk or DC voltage 170 V Max DC or AC pk carry current: 1A Max DC or AC pk switching current: 0.5A Max. contact rating: 10 W |
Signal Bandwidth | |
Input to Output (input switch card to LF switch card output) | > 20 MHz |
HF Switch Card (external input to output) | > 100 MHz |
I/O Connections | |
GX6032 Switch Card | 32 coaxial connections (Positronics BD8W8F85R70T0/AA) |
GX6256 Switch Card | Receiver: Dual 200 pin connector Digital I/O: 34 pin, dual row latched header |
GX6192 Switch Card | Receiver: Dual 200 pin connector Digital I/O: 34 pin, dual row latched header External I/O: 16 coaxial connections (Positronics BD8W8F85R70T0/AA) |
Dynamic Digital Subsystem | |
---|---|
Configuration | 6U PXI format, (7) 32 channel DIO cards, (1) 16 channel DIO card, programmable drive / sense per pin, 50 MHz vector rate (max) |
Number of Hyrid Channels | 224 |
Test Modes | Dynamic or static |
Data Output Formats (per channel) | Drive Hi, Drive Lo, Hi-Z Formatted Data: No return, Return to 1, Return to 0, Return to Hi-Z, Return to complement, Surround by complement; selectable on a per channel basis |
Drive Data Timing (per channel) | Data assert / de-assert based on Phases 1-4 |
Timing Sets | 256 Timing Set groups with 4 Phases, 4 Windows, and 4K sequence steps) |
Phase Programming Range (Assert/Return) | 0 ns to 64 us (using the 500 MHz master clock) |
Window Programming Range (Open/Close) | 0 ns to 64 us (using the 500 MHz master clock) |
Phase and Window Timing Resolution | 1 ns, using the 500 MHz master clock |
Capture Modes (per channel) | Mask Opening edge of Window Closing edge of Window Window – data is valid for entire window duration |
Drive/Expect Mode | Output: Drive Hi, Drive Lo, Hi-Z Expect: 1, 0, OK, between states, or mask Keep last Toggle last Accumulate CRC-16 |
Error Address Record | Record address for memory errors 1K deep error memory |
Drive Level Range | -14 V to +26 V |
Drive Voltage Level Range | Min: 0.5 V p-p Max: 26 V p-p |
Drive Voltage Accuracy | ± 25 mv, < 26 V p-p drive voltage |
Sense Level Range | -16 V to +22 V |
Maximum Sense Voltage Range | 26 V |
Sense Voltage Resolution | 16 bits |
Pull-Up/Pull-Down Current Source/Sink | ± 24 ma, programmable on a per channel basis V commutate: -16 to +22 V., programmable on a per channel basis |
Pull-Up/Pull-Down Current Source/Sink Accuracy | ± 250 uA |
Pull-Up/Pull-Down Current Source/Sink Resolution | 16 bits |
Trigger/Clock Signals | (4) Aux signals for external triggering, clocking of digital subsystem, accessible via the receiver interface and the IIP |
Mainframe Specifications | |
Input AC Power | 120 / 208 VAC nominal, 3 phase, 5-wire wye configuration 20 A per phase, 47 to 63 Hz, Input voltage range: 100 / 170 to 155 / 264 VAC |
HV Input AC Power Configuration | 230 / 400 VAC nominal, 3 phase, 5-wire wye configuration 10 A per phase, 47 to 63 Hz, input voltage range: 223 / 380 to 310 / 528 VAC |
Total Available DC Power | PXI voltages: 800 W Pin electronics: 4.4 KW |
PXI Power | |
+5 V +3.3 V +12 V -12 V Note: Total output power cannot exceed 800 W. |
60 A (max) 60 A (max) 25 A (max) 5 A (max) |
Pin Electronics (via P5 connector) | |
VCC VEE 3.3 V |
+2 V to +30 V, @ 50 A -4 V to -18 V, @ 80A 3.3 V @ 60 A |
Power Supply Monitoring | Monitored voltages: 3.3, 5, +12, -12, VIO value Accuracy: ± 2% of reading |
Temperature Monitoring | Per slot monitoring, 1 reading/sec/slot 4 second moving average value User selectable alarm criteria:
Accuracy: ± 2 °C |
PXI Triggers | Slots: 2 – 20 Number: 8 per segment Software controlled segment mapping supports:
|
PXI Clock | Integrated 10 MHz PXI clock with auto-detect function. Presence of an external 10 MHz PXI clock will disable the internal clock. PXI clock is distributed to all peripheral slots.
|
Slots | 20 PXI or cPCI Slots (19 instruments max) |
Weight (core system) | 100 lbs |
Overall Dimensions | 10U (17.5″) H x 19″ W x 24″ D |
Cooling | Eight 100 CFM fans for system cooling. Integrated temperature monitoring via an on-board microcontroller with audible and software notification when preset temperature limits are exceeded. Fan speed control and monitoring is automatic and can be controlled / monitored via the GxChassis software. |
Environmental and Compliance | |
Operating | 0 °C to +50 °C |
Storage | -20 °C to +60 °C |
Relative Humidity (operating) | 5% to 80% RH,non-condensing |
Relative Humidity (non-operating) | 5% to 95% RH,non-condensing |
Altitude (operating) | Up to 2000 M |
CE Compliance | EN61010-1 (pending) EN61326 (pending) |
Note: Specifications are subject to change without notice.