GX1838 Series Specifications
Number of Output Channels | 8 |
Number of Voltage Rails | 3 |
Number of DACs | 3 |
Output Voltage Ranges | -10 V to +32 VDC (all rails) -20 V to +20 VDC (all rails) |
Resolution | 14-bit |
Accuracy @ 1 kOhm Load |
±14 mV |
Slew Rate | 6 V/µs |
Maximum Load (Minimum Resistance) | 50 Ω |
Output Noise | < 4 mV rms, 1 KΩ load, 20 Hz – 300 KHz measurement BW |
Output Protection | Short Circuit (Ground) |
Power ON State | All Channels / Rails open |
Maximum Current * | |
---|---|
Per Channel | 500 mA |
Per Rail | 500 mA |
Per GX1838 | 500 mA |
PXI Power Requirements | |
+5 V | 2.9 A, max |
+3.3 V | 2.1 A max |
+12 V | 0.55 A, max |
-12 V | 0.55 A, max |
Environmental | |
Operating Temperature | 0 °C to +55 °C |
Storage Temperature | -20 °C to +70 °C |
Size | 3U PXI |
Weight | 14 oz |
Calibration | |
Calibration Interval | 1 year |
*The total maximum current is 500 mA (i.e., if one channel or rail uses the entire 500 mA, the other channels or rails cannot be used). Note: Specifications are subject to change without notice.