Arbitrary Waveform Function Generator PXI Card

  • Arbitrary Waveform Generator and Direct Digital Synthesis modes
  • 100 MS / s sample rate
  • 12–bit vertical resolution
  • Programmable from 100 mVPP to 8 VPP, full scale
  • 2 MS memory
  • PLL clock generator for AWG mode
  • 3U, single slot module
Send Enquiry

GX1110 Specifications

Standard Waveforms
Sine, triangle, square, pulse, ramp up, ramp down, sinc (sine(x) / x), Guassian pulse, noise
Maximum Frequency
(FS = 100 MHz)
Waveform DDS Mode AWG Mode
Sine 30 MHz 6.25 MHz
Triangle 1 MHz 6.25 MHz
Ramp 1 MHz 6.25 MHz
Pulse 6.25 MHz
Square 30 MHz 50 MHz
Noise 1 MHz
Gaussian pulse 1 MHz
Sin(x) / x 1 MHz
Arbitrary Waveform Generator Mode
Sample Rate 10 mS/s to 100 MS/s
Sample Rate Resolution and Accuracy 4 digits or 0.01 Hz, which ever is lower
0.01% accuracy
Vertical Resolution 12 bits
Waveform Memory 2 M samples
DDS Mode
Frequency Range,
Resolution, and
Accuracy
10 µHz to 30 MHz
10 µHz, resolution
50 ppm accuracy (OCXO option available for higher accuracy)
Non-Harmonic Spurious Components < 60 dBc (DC to 1 MHz)
Distortion
(2nd Harmonic Relative to Carrier)
< -60 dBc @ < 200 kHz
< -35 dBc @ < 2.0 MHz
< -20 dBc @ < 20 MHz
Modulation AM: Int. / Ext. 0 – 100%, DC to 20 kHz
FM: Int. / Ext. DC to 20 kHz
Phase offset: 0 – 360 degrees, 0.1 degree resolution
Main Output
Connector Front panel BNC
Output Mode On / Off
Output Impedance 50 Ω, ±1%
Protection Protected against shorts to ground and over-voltage
Amplitude Range 100 mVPP to 8 VPP into 50 Ω; double into open circuit. 0.1, 0.3, 1, 3, and 8 VPP, FS ranges.
Amplitude Resolution 3 digits
Amplitude Accuracy ±(2% of programmed value + 5 mV) < 1 V
±(1% of programmed value + 10 mV) ≥ 1 V
Offset Range 0 to ±5 V, amplitude dependent
Offset Resolution 3 digits
Offset Accuracy ±(1% of programmed value + 20 mV)
Rise Time < 15 ns into a 50 Ω load (10% to 90% full scale step)
Abberations < 5% of p-p amplitude, ±50 mV
Flatness ±.5 dB to 10 MHz, ±1 dB to 30 MHz
Filters 30 MHz, elliptical; DDS mode
30 MHz, bessel; AWG mode
SYNC Output
Connector Front panel BNC output is synchronous with output waveform
Impedance 50 Ω, ±1%
Level TTL compatible
Marker Output
Positive pulse, user programmable and synchronous with the waveform output.
Connector Front panel PS2
Impedance 50 Ω, ±1%
Level TTL compatible
Trigger Modes
Continuous Output continuously generated
Triggered Output waveform triggered by external or software triggered event. The external trigger signal edge may be a rising or falling edge.
Start / Stop Output waveform triggered by a trigger signal edge. Waveform is continuously generated until the occurrence of an another trigger edge. The trigger signal may be external or software generated. The external trigger signal edge may be the rising or falling edges.
Gated Same as Triggered mode except that the waveform is generated for as long as the gate signal stays true (logic one). When the gate signal goes false (logic zero) the output goes quiescent.
Gated Hold Same as Gated mode except that when the gate signal goes true (logic high) the waveform freezes until the gate signal goes low.
Burst Output waveform will become active on the occurrence of a trigger edge. The selected waveform is generated for a preset number of cycles between 2 and 1,048,576. Output will then disable.
External Trigger Input
Connector Front panel PS2
Impedance 10 K Ω nominal
Threshold Level TTL
Repetition Rate DC to 10 MHz
Minimum Pulse Width 50 ns
Slope Positive or negative going edge
External Input / Output Clock
Connector Front panel BNC
Input Impedance 10 K Ω nominal
Threshold Level TTL
Output Impedance 50 Ω
External Modulation Input
Connector Front panel PS2
Input Voltage Range 5 VPP for 100% modulation (AM & FM)
Input Impedance 10 K Ω nominal
Bandwidth DC to 20 kHz
General
Power Requirements 15 W (max)
Current Consumption (Maximum) +5 V @ 1 A
+12 V @ 300 mA
-12 V @ 300 mA
+3.3 V @ 1 A
Weight Approx. 12 oz
Size 3U, single slot
Operating Temperature 0 °C to +50 °C
Storage Temperature -20 °C to +70 °C
Humidity
(Non-Condensing)
5% to 95%, < 70 °C
Safety Designed to meet IEC 1010-1, UL 3111-1, and CSA 22.2#1010
CE Labeled Yes
Calibration Interval 1 year

Note: Specifications are subject to change without notice.

The GX1110 is a high performance, single-channel PXI arbitrary waveform generator that offers function generator and arbitrary waveform generator functionality within one instrument. Built-in waveforms are available for use with both the DDS or AWG modes of operation and include Sine, Triangle, Ramp, Noise, Gaussian pulse and Sinx / x. A flexible sequencer is also available as part of the AWG’s architecture, supporting the generation of complex waveforms.

Features

With a sample rate of 100 MS/s, the GX1110 is an ideal modulation source for troubleshooting encoding schemes as well as simulating signal distortion, power line cycle dropouts, video signals, components failures and power supply transients.

The GX1110 Series comes standard with 2 M samples of waveform memory, accessible via a high speed interface.

Triggering
The GX1110 series provides an external trigger input, offering users the ability to control waveform generation via external trigger events. When operating in the AWG mode, waveform trigger modes include continuous, triggered, gated hold, and burst functionality.

Sample Clock
The GX1110 Series utilizes a PLL, providing a programmable sample clock for the AWG mode of operation. The reference clock is provided by an on-board crystal oscillator, or alternately, an external 10 MHz reference can be used. When operating in the DDS mode, an on-board crystal oscillator source provides a 100 MHz clock source for the DDS generator and associated D to A converter.

Programming and Software

The board is supplied with the GtWave, a software package that includes a virtual instrument panel, a Windows 32/64-bit DLL driver library and documentation. A trial license for WaveEasy, a software to create, modify and analyze waveform is also included. The The virtual panel can be used to interactively program and control the instrument from a window that displays the instrument’s current settings and status. In addition, interface files are provided to support access to programming tools and languages such as ATEasy, LabView, LabView/Real-Time, C/C++, Microsoft Visual Basic®, Delphi, and Pascal. An On-Line help file and PDF User’s Guide provides documentation that includes instructions for installing, using and programming the board.

Applications

  • Video
  • Navigation
  • Radar
  • Sonar
  • Communications
  • Converter Testing
  • Filter Design & Test
  • Computer Peripherals

Enjoy:

Send Enquiry For:

GX1110

Israel's Leading Businesses Choose DCT

Israel's Leading Businesses Choose DCT

DCT’s Skilled Management Team has more than twenty-five years of experience and expertise in the local business environment, coupled with a highly motivated group of employees, positions us as the preferred supplier among the major customers in the industry.

DCT Competitive Edge is the prompt response to our customer needs, the online and on-site direct technical support, professional consulting for the best cost/performance solution, on-time delivery, long-term maintenance, and support.

We can get back to you!

Need a Rugged Device?

Fill in the form!

A variety of tablets are available for all use cases.
- only solutions - No obligations -

Our experts can help you find just about anything.

What Were You Looking For? Let us help.