| Number of input channels |
4 analog |
| CH1, CH2, CH3, CH4 |
Isolated female BNC |
| Type |
Differential input |
| Resolution |
User selectable in software |
| Native |
8, 12, 14Â bit |
| Enhanced |
16Â bit |
| DC Accuracy |
0.25 % (0.1 % typical) of full scale ± 1 LSB at 20°C to 25°C
To achieve rated accuracy, allow the instrument to settle for 20 minutes. When subjected to extreme temperatures, allow additional time for internal temperatures to stabilize. |
| Noise |
305 µVrms (200 mV range, 12 bit, 50 MSa/s)
90 µVrms (200 mV range, 16 bit, 6.25 MSa/s) |
| Ranges (full scale) |
±200 mV
±400 mV
±800 mV |
±2 V
±4 V
±8 V |
±20 V
±40 V
±80 V |
| Coupling |
AC/DC |
| Impedance |
2 MOhm / 12 pF ± 1 %
1 MOhm / 20 pF ± 1 % when SafeGround enabled |
| Maximum voltage |
200Â V (DC + AC peak < 10Â kHz) |
| Maximum Common Mode voltage |
200Â mV to 800Â mV ranges : 2Â V
2Â V to 8Â V ranges : 20Â V
20Â V to 80Â V ranges : 200Â V |
| Common Mode Rejection Ratio |
-47Â dB |
| Bandwidth |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
| -3dBÂ at 75 % of full scale input |
250Â MHz |
250Â MHz |
250Â MHz |
100Â MHz |
100Â MHz |
| Limit, selectable per channel |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (100Â MHz)
75Â MHz
50Â MHz
25Â MHz |
Off (100Â MHz)
75Â MHz
50Â MHz
25Â MHz |
| AC coupling cut off frequency (-3dB) |
±1.5 Hz |
| Channel Isolation |
500Â V |
| Channel Separation |
-80Â dB |
| SureConnect |
Optionally available (option S) |
| Maximum voltage on connection |
200Â V (DC + AC peak < 10Â kHz) |
| Resistance measurement |
Optionally available (option S) |
| Ranges (Full scale) |
100 Ω
200 Ω
500 Ω |
1 kΩ
2 kΩ
5 kΩ |
10 kΩ
20 kΩ
50 kΩ |
100 kΩ
200 kΩ
500 kΩ |
1 MΩ
2 MΩ |
| Accuracy |
1 % of full scale |
| Response time (to 95%) |
< 10 µs |
| SafeGround |
Optionally available (option G) |
| Maximum voltage on connection |
200Â V (DC + AC peak < 10Â kHz) |
| Maximum switch off current |
500Â mA |
| Response time |
< 100Â ns |
|
| Maximum sampling rate |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
| 8Â bit |
|
| measuring one channel |
1 GSa/s |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
| measuring two channels |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
| measuring three or four channels |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
| 12Â bit |
|
| measuring one channel |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
| measuring two channels |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
| measuring three or four channels |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
5 MSa/s |
| 14Â bit |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
5 MSa/s |
| 16Â bit |
6.25 MSa/s |
3.125 MSa/s |
1.25 MSa/s |
625 kSa/s |
312.5 kSa/s |
|
| Maximum streaming rate 1 |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
| 8Â bit |
|
| measuring one channel |
200 MSa/s2 |
100 MSa/s2 |
40 MSa/s |
20 MSa/s |
10 MSa/s |
| measuring two channels |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
| measuring three or four channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
| 12Â bit |
|
| measuring one channel |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
| measuring two channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
| measuring three or four channels |
25 MSa/s5 |
12.5 MSa/s5 |
5 MSa/s |
2.5 MSa/s |
1.25 MSa/s |
| 14Â bit |
|
| measuring one channel |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
| measuring two channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
| measuring three or four channels |
25 MSa/s5 |
12.5 MSa/s5 |
5 MSa/s |
2.5 MSa/s |
1.25 MSa/s |
| 16Â bit |
6.25 MSa/s6 |
3.125 MSa/s |
1.25 MSa/s |
625 kSa/s |
312.5 kSa/s |
- On some computers, the highest streaming rates may not be available, due to computer restrictions.
- ≤ 40 MSa/s when connected to USB 2.0
- ≤ 20 MSa/s when connected to USB 2.0
- ≤ 10 MSa/s when connected to USB 2.0
- ≤ 5 MSa/s when connected to USB 2.0
- ≤ 3.125 MSa/s when connected to USB 2.0
|
| Sampling clock source |
|
| Internal |
TCXO |
| Accuracy |
± 0.0001 % |
| Stability |
± 1 ppm over 0°C to 55°C |
| Time base aging |
±1 ppm/year |
| External |
LVDS, on CMI connectors |
| Input frequency |
10 MHz ± 1 %
16.369 MHz ± 1 % |
|
| Memory |
Standard model |
XMÂ option |
| 8Â bit resolution |
|
| Measuring one channel |
1 Mpts per channel |
256 Mpts per channel |
| Measuring two channels |
512 Kpts per channel |
128 Mpts per channel |
| Measuring three or four channels |
256 Kpts per channel |
64 Mpts per channel |
| 12, 14, 16Â bit resolution |
|
| Measuring one channel |
512 Kpts per channel |
128 Mpts per channel |
| Measuring two channels |
256 Kpts per channel |
64 Mpts per channel |
| Measuring three or four channels |
128 Kpts per channel |
32 Mpts per channel |
|
| Trigger |
Only available in scope/block mode, not in streaming mode |
| System |
Digital, 2 levels |
| Source |
CH1, CH2, CH3, CH4, OR, digital external |
| Trigger modes |
Rising edge, Falling edge, Any edge
Inside window, Outside window, Enter window, Exit window (time conditions only on Ch1)
Pulse width (only on Ch1)
Interval (only on Ch1) |
| Level adjustment |
0 to 100 % of full scale |
| Hysteresis adjustment |
0 to 100 % of full scale |
| Resolution |
0.006 % (14, 16Â bits) / 0.025% (12Â bits) |
| Pre trigger |
0 to selected record length, 1 sample resolution |
| Post trigger |
0 to selected record length, 1 sample resolution |
| Trigger hold-off |
0 to 63Â MSamples, 1 sample resolution |
| Trigger delay |
0 to 8Â GSamples, 1 sample resolution |
| Digital external trigger |
|
| Input |
Extension connector pins 1 and 2 |
| Range |
0 to 2.5Â V (TTL) |
| Coupling |
DC |
| Jitter |
≤ 1 sample |
| Using CMI |
Synchronizing instruments via CMI is available when all instruments are connected via USB. |
| Maximum number of instruments |
Limited by number of available USB ports |
| Synchronization accuracy |
0 ppm |
| CMI interface |
2x, CMI 1, CMI 2 |
| Required coupling cable |
TP-C50H Coupling cable CMI |
| Maximum coupling cable length |
50Â cm |
| Using WCMI |
Synchronizing instruments via WCMI is available when instruments are connected via USB, LAN and WiFi. |
| Maximum number of instruments |
No limitation |
| Required coupling module |
WCMI – 8, WCMI – 9 |
| Clock synchronization accuracy |
≤ 1 ppm, typical ≤ 0.2 ppm |
| Trigger jitter at sample rate 1 |
≤ 1 MSa/s |
> 1 MSa/s |
|
|
| 2 x “5” |
≤ ±2 samples |
≤ ±2 µs |
|
|
| “5” and “6” |
|
| Trigger source “5” |
≤ ±2 samples |
≤ ±2 µs |
|
|
| Trigger source “6” |
≤ ±8 samples |
≤ ±8 µs |
|
|
| 2 x “6” |
≤ ±8 samples |
≤ ±8 µs |
|
|
- “5” = WiFiScope WS5 or Handyscope HS5. “6” = WiFiScope WS6 DIFF, WiFiScope WS6, Handyscope HS6 DIFF, Handyscope HS6, ATS610004DW-XMSG, ATS610004D-XMSG, ATS605004DW-XMS or ATS605004D-XMS
|
| Output |
Extension connector pins 3 (signal) and 6 (ground) |
| Signal |
Square wave |
| Level |
-1Â V to 1Â V |
| Frequency |
1Â kHz |
| Interface |
USB 3.0 SuperSpeed (5Â Gbit/s);
(USB 2.0 HighSpeed compatible) |
| Power |
From USB port or external input |
| Consumption |
5Â VDCÂ 1200Â mA max |
| External power |
From second USB port or power adapter |
| Instrument |
|
| Height |
25Â mm (1Â inch) |
| Length |
170Â mm (6.7Â inch) |
| Width |
140Â mm (5.2Â inch) |
| Weight |
500Â g (17.6Â ounce) |
| USB cord length |
1.8Â m (70Â inch) |
| Channel 1, 2, 3, 4 |
Isolated BNC, female |
| USB |
Fixed cable with USB 3.0 type A connector, 1.8Â m |
| Extension connector |
D-sub 9 pins female |
| Power |
3.5Â mm power socket |
| CMI I/O connectors |
2 x HDMI type C socket |
| PC I/O connection |
USB 2.0, USB 3.0 or USB 3.1 |
| Operating system |
Windows 10 / 11, 64Â bits
Linux |
| Operating |
|
| Ambient temperature |
20°C to 25°C (10°C to 40°C without specifications) |
| Relative humidity |
10 % to 90 %, non condensing |
| Storage |
|
| Ambient temperature |
-20°C to 70°C |
| Relative humidity |
5 % to 95 %, non condensing |
| CE mark compliance |
Yes |
| RoHS |
Yes |
| EN 55011:2009/A1:2010 |
Yes |
| EN 55022:2006/A1:2007 |
Yes |
| EN 61000-6-1:2007 |
Yes |
| EN 61000-6-3:2007 |
Yes |
| Instrument |
Handyscope HS6 DIFF |
| Measure leads |
4 x Measure lead TP-C812B, Differential BNC -> 4Â mm banana plug |
| Differential attenuator |
4 x Differential attenuator TP-DA10 |
| Alligator clips, large |
Alligator Clip TP-AC80I – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 large alligator clips |
| Alligator clips, medium |
Alligator Clip TP-AC10I – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 medium alligator clips |
| Alligator clips, small |
Alligator Clip TP-AC5 – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 small alligator clips |
| Accessories |
Handyscope / WiFiScope power supply, external power cable for USB port, Probe compensation cable TP-DB9-BNC-30 (only with option SafeGround) |
| Software |
for Windows 10 / 11, 64 bits |
| Drivers |
for Windows 10 / 11, 64 bits |
| Software Development Kit |
for Windows 10 / 11, 64 bits, and Linux |
| Manuals |
instrument manual and software user’s manuals
color printed and digital |