Number of input channels |
4 analog |
CH1, CH2, CH3, CH4 |
Isolated female BNC |
Type |
Differential input |
Resolution |
User selectable in software |
Native |
8, 12, 14Â bit |
Enhanced |
16Â bit |
DC Accuracy |
0.25 % (0.1 % typical) of full scale ± 1 LSB at 20°C to 25°C
To achieve rated accuracy, allow the instrument to settle for 20 minutes. When subjected to extreme temperatures, allow additional time for internal temperatures to stabilize. |
Noise |
305 µVrms (200 mV range, 12 bit, 50 MSa/s)
90 µVrms (200 mV range, 16 bit, 6.25 MSa/s) |
Ranges (full scale) |
±200 mV
±400 mV
±800 mV |
±2 V
±4 V
±8 V |
±20 V
±40 V
±80 V |
Coupling |
AC/DC |
Impedance |
2 MOhm / 12 pF ± 1 %
1 MOhm / 20 pF ± 1 % when SafeGround enabled |
Maximum voltage |
200Â V (DC + AC peak < 10Â kHz) |
Maximum Common Mode voltage |
200Â mV to 800Â mV ranges : 2Â V
2Â V to 8Â V ranges : 20Â V
20Â V to 80Â V ranges : 200Â V |
Common Mode Rejection Ratio |
-47Â dB |
Bandwidth |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
-3dBÂ at 75 % of full scale input |
250Â MHz |
250Â MHz |
250Â MHz |
100Â MHz |
100Â MHz |
Limit, selectable per channel |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (250Â MHz)
150Â MHz
100Â MHz
50Â MHz |
Off (100Â MHz)
75Â MHz
50Â MHz
25Â MHz |
Off (100Â MHz)
75Â MHz
50Â MHz
25Â MHz |
AC coupling cut off frequency (-3dB) |
±1.5 Hz |
Channel Isolation |
500Â V |
Channel Separation |
-80Â dB |
SureConnect |
Optionally available (option S) |
Maximum voltage on connection |
200Â V (DC + AC peak < 10Â kHz) |
Resistance measurement |
Optionally available (option S) |
Ranges (Full scale) |
100 Ω
200 Ω
500 Ω |
1 kΩ
2 kΩ
5 kΩ |
10 kΩ
20 kΩ
50 kΩ |
100 kΩ
200 kΩ
500 kΩ |
1 MΩ
2 MΩ |
Accuracy |
1 % of full scale |
Response time (to 95%) |
< 10 µs |
SafeGround |
Optionally available (option G) |
Maximum voltage on connection |
200Â V (DC + AC peak < 10Â kHz) |
Maximum switch off current |
500Â mA |
Response time |
< 100Â ns |
|
Maximum sampling rate |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
8Â bit |
|
measuring one channel |
1 GSa/s |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
measuring two channels |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
measuring three or four channels |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
12Â bit |
|
measuring one channel |
500 MSa/s |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
measuring two channels |
200 MSa/s |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
measuring three or four channels |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
5 MSa/s |
14Â bit |
100 MSa/s |
50 MSa/s |
20 MSa/s |
10 MSa/s |
5 MSa/s |
16Â bit |
6.25 MSa/s |
3.125 MSa/s |
1.25 MSa/s |
625 kSa/s |
312.5 kSa/s |
|
Maximum streaming rate 1 |
HS6D-1000 |
HS6D-500 |
HS6D-200 |
HS6D-100 |
HS6D-50 |
8Â bit |
|
measuring one channel |
200 MSa/s2 |
100 MSa/s2 |
40 MSa/s |
20 MSa/s |
10 MSa/s |
measuring two channels |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
measuring three or four channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
12Â bit |
|
measuring one channel |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
measuring two channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
measuring three or four channels |
25 MSa/s5 |
12.5 MSa/s5 |
5 MSa/s |
2.5 MSa/s |
1.25 MSa/s |
14Â bit |
|
measuring one channel |
100 MSa/s3 |
50 MSa/s3 |
20 MSa/s |
10 MSa/s |
5 MSa/s |
measuring two channels |
50 MSa/s4 |
25 MSa/s4 |
10 MSa/s |
5 MSa/s |
2.5 MSa/s |
measuring three or four channels |
25 MSa/s5 |
12.5 MSa/s5 |
5 MSa/s |
2.5 MSa/s |
1.25 MSa/s |
16Â bit |
6.25 MSa/s6 |
3.125 MSa/s |
1.25 MSa/s |
625 kSa/s |
312.5 kSa/s |
- On some computers, the highest streaming rates may not be available, due to computer restrictions.
- ≤ 40 MSa/s when connected to USB 2.0
- ≤ 20 MSa/s when connected to USB 2.0
- ≤ 10 MSa/s when connected to USB 2.0
- ≤ 5 MSa/s when connected to USB 2.0
- ≤ 3.125 MSa/s when connected to USB 2.0
|
Sampling clock source |
|
Internal |
TCXO |
Accuracy |
± 0.0001 % |
Stability |
± 1 ppm over 0°C to 55°C |
Time base aging |
±1 ppm/year |
External |
LVDS, on CMI connectors |
Input frequency |
10 MHz ± 1 %
16.369 MHz ± 1 % |
|
Memory |
Standard model |
XMÂ option |
8Â bit resolution |
|
Measuring one channel |
1 Mpts per channel |
256 Mpts per channel |
Measuring two channels |
512 Kpts per channel |
128 Mpts per channel |
Measuring three or four channels |
256 Kpts per channel |
64 Mpts per channel |
12, 14, 16Â bit resolution |
|
Measuring one channel |
512 Kpts per channel |
128 Mpts per channel |
Measuring two channels |
256 Kpts per channel |
64 Mpts per channel |
Measuring three or four channels |
128 Kpts per channel |
32 Mpts per channel |
|
Trigger |
Only available in scope/block mode, not in streaming mode |
System |
Digital, 2 levels |
Source |
CH1, CH2, CH3, CH4, OR, digital external |
Trigger modes |
Rising edge, Falling edge, Any edge
Inside window, Outside window, Enter window, Exit window (time conditions only on Ch1)
Pulse width (only on Ch1)
Interval (only on Ch1) |
Level adjustment |
0 to 100 % of full scale |
Hysteresis adjustment |
0 to 100 % of full scale |
Resolution |
0.006 % (14, 16Â bits) / 0.025% (12Â bits) |
Pre trigger |
0 to selected record length, 1 sample resolution |
Post trigger |
0 to selected record length, 1 sample resolution |
Trigger hold-off |
0 to 63Â MSamples, 1 sample resolution |
Trigger delay |
0 to 8Â GSamples, 1 sample resolution |
Digital external trigger |
|
Input |
Extension connector pins 1 and 2 |
Range |
0 to 2.5Â V (TTL) |
Coupling |
DC |
Jitter |
≤ 1 sample |
Using CMI |
Synchronizing instruments via CMI is available when all instruments are connected via USB. |
Maximum number of instruments |
Limited by number of available USB ports |
Synchronization accuracy |
0 ppm |
CMI interface |
2x, CMI 1, CMI 2 |
Required coupling cable |
TP-C50H Coupling cable CMI |
Maximum coupling cable length |
50Â cm |
Using WCMI |
Synchronizing instruments via WCMI is available when instruments are connected via USB, LAN and WiFi. |
Maximum number of instruments |
No limitation |
Required coupling module |
WCMI – 8, WCMI – 9 |
Clock synchronization accuracy |
≤ 1 ppm, typical ≤ 0.2 ppm |
Trigger jitter at sample rate 1 |
≤ 1 MSa/s |
> 1 MSa/s |
|
|
2 x “5” |
≤ ±2 samples |
≤ ±2 µs |
|
|
“5” and “6” |
|
Trigger source “5” |
≤ ±2 samples |
≤ ±2 µs |
|
|
Trigger source “6” |
≤ ±8 samples |
≤ ±8 µs |
|
|
2 x “6” |
≤ ±8 samples |
≤ ±8 µs |
|
|
- “5” = WiFiScope WS5 or Handyscope HS5. “6” = WiFiScope WS6 DIFF, WiFiScope WS6, Handyscope HS6 DIFF, Handyscope HS6, ATS610004DW-XMSG, ATS610004D-XMSG, ATS605004DW-XMS or ATS605004D-XMS
|
Output |
Extension connector pins 3 (signal) and 6 (ground) |
Signal |
Square wave |
Level |
-1Â V to 1Â V |
Frequency |
1Â kHz |
Interface |
USB 3.0 SuperSpeed (5Â Gbit/s);
(USB 2.0 HighSpeed compatible) |
Power |
From USB port or external input |
Consumption |
5Â VDCÂ 1200Â mA max |
External power |
From second USB port or power adapter |
Instrument |
|
Height |
25Â mm (1Â inch) |
Length |
170Â mm (6.7Â inch) |
Width |
140Â mm (5.2Â inch) |
Weight |
500Â g (17.6Â ounce) |
USB cord length |
1.8Â m (70Â inch) |
Channel 1, 2, 3, 4 |
Isolated BNC, female |
USB |
Fixed cable with USB 3.0 type A connector, 1.8Â m |
Extension connector |
D-sub 9 pins female |
Power |
3.5Â mm power socket |
CMI I/O connectors |
2 x HDMI type C socket |
PC I/O connection |
USB 2.0, USB 3.0 or USB 3.1 |
Operating system |
Windows 10 / 11, 64Â bits
Linux |
Operating |
|
Ambient temperature |
20°C to 25°C (10°C to 40°C without specifications) |
Relative humidity |
10 % to 90 %, non condensing |
Storage |
|
Ambient temperature |
-20°C to 70°C |
Relative humidity |
5 % to 95 %, non condensing |
CE mark compliance |
Yes |
RoHS |
Yes |
EN 55011:2009/A1:2010 |
Yes |
EN 55022:2006/A1:2007 |
Yes |
EN 61000-6-1:2007 |
Yes |
EN 61000-6-3:2007 |
Yes |
Instrument |
Handyscope HS6 DIFF |
Measure leads |
4 x Measure lead TP-C812B, Differential BNC -> 4Â mm banana plug |
Differential attenuator |
4 x Differential attenuator TP-DA10 |
Alligator clips, large |
Alligator Clip TP-AC80I – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 large alligator clips |
Alligator clips, medium |
Alligator Clip TP-AC10I – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 medium alligator clips |
Alligator clips, small |
Alligator Clip TP-AC5 – Set of 8: Green, Blue, Yellow, Red, 4 Black, 8 small alligator clips |
Accessories |
Handyscope / WiFiScope power supply, external power cable for USB port, Probe compensation cable TP-DB9-BNC-30 (only with option SafeGround) |
Software |
for Windows 10 / 11, 64 bits |
Drivers |
for Windows 10 / 11, 64 bits |
Software Development Kit |
for Windows 10 / 11, 64 bits, and Linux |
Manuals |
instrument manual and software user’s manuals
color printed and digital |