GX1034 Specifications
| General | |
|---|---|
| Source / Measure Channels | Meas Hi, Meas Lo Sense Hi, Sense Lo |
| Source / Measure Connector | DB-9, female |
| 10 MHz Reference Connector | BNC front panel and PXI backplane connector |
| Format | PXI, 3U single slot, compatible with PXI Express hybrid slots |
| DC Voltage Source Standard | |
| Output Voltage | ±9.0, ±0.9, ±0.09, ±0.009 V |
| Accuracy | ±9.0 V: ±0.005% ±0.9 V: ±0.005% ±0.09 V: ±0.016% ±0.009 V: ±0.1% |
| Temperature Coefficent | 0.6 ppm / °C |
| Long Term Drift | 6 ppm for 1000 hours |
| Output Divider | Resistive series network (Ohms): 10 k, 1 k, 100, 10 |
| AC Voltage Source Standard | |
| Output Voltage | 3.5, 0.35, 0.035, 0.0035 VAC, nominal All voltages support the following frequencies: 50 Hz, 60 Hz, 100 Hz, 400 Hz, 1 kHz, 10 kHz, 100 kHz |
| Accuracy 50 Hz, 60 Hz, 100 Hz |
3.5 VAC: ±0.6% 0.35 VAC: ±0.6% 0.035 VAC: ±0.6% 0.0035 VAC: ±4% |
| Accuracy 400 Hz, 1 kHz |
3.5 VAC: ±0.6% 0.35 VAC: ±0.6% 0.035 VAC: ±0.6% 0.0035 VAC: ±4.75% |
| Accuracy 10 kHz |
3.5 VAC: ±0.6% 0.35 VAC: ±0.6% 0.035 VAC: ±0.6% 0.0035 VAC: ±5% |
| Accuracy 100 kHz |
3.5 VAC: ±4% 0.35 VAC: ±4% 0.035 VAC: ±4% 0.0035 VAC: ±9.5% |
| Temperature Coefficient | 3 ppm / °C (voltage reference) |
| Long Term Drift | 6 ppm for 1000 operating hours |
| Frequency Accuracy | ±0.1 ppm |
| Output Divider | Resistive series network (Ohms): 10 k, 1 k, 100, 10 |
| Resistor Standards | |
|---|---|
| 1 Ohm | ±0.06%, 3 ppm / °C, 4-wire |
| 10 Ohm | ±0.012%, 0.2 ppm / °C, 4-wire |
| 100 Ohm | ±0.01%, 0.2 ppm / °C, 4-wire |
| 1 kOhm | ±0.004%, 0.2 ppm / °C, 2-wire |
| 10 kOhm | ±0.004%, 0.2 ppm / °C, 2-wire |
| 100 kOhm | ±0.015%, 5 ppm / °C, 2-wire |
| 1 MOhm | ±0.0275%, 3 ppm / °C, 2-wire |
| 10 MOhm | ±0.1%, 2 ppm / °C, 2-wire |
| 10 MHz Frequency Standard and Divider | |
| Output Frequency | 10 MHz |
| Initial Accuracy | ±1 ppm |
| Drift | 100 ppb / °C (max) |
| Output | Front panel BNC: TTL compatible, 50 Ohm load PXI connector: PXI 10 MHz clock |
| 10 MHz Divider Output | 24-bit resolution, programmable |
| DC Current Source | |
| Range | 0 – 19.9 mA 16-bit resolution, programmable |
| Accuracy | ±0.025 mA of programmed value |
| Compliance | 0 V – 12.5 V |
| DC Voltage Measurement | |
| Resolution | 16-bit |
| Mode | Differential |
| Input Impedance | 20 kOhm nominal |
| Range | -10 V to +10 V |
| Accuracy | ±0.025 V of reading |
| Power | |
| +5 V | 1 A (max.) |
| +12 V | 500 mA (max.) |
| +3.3 V | 500 mA (max.) |
| Configuration | |
| Weight | 210 g |
| Size | 3U, single slot, hybrid slot compatible |
| Environmental | |
| Operating Temperature | 0 °C to +50 °C |
| Storage Temperature | 0 °C to +70 °C |
| Humidity | 10% to 80% (Non-condensing) |
| Safety | EN61010-1:2001 |
| CE Labeled | Yes EN61000-6-1:2001, EN55011:1998 |
| Calibration Interval | 1 year |